Aoralscan 3_Manual_V1.0.0.30 - 15 -
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic environment
–
guidance
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
(100% dip in UT)
for 0.5/1 cycle
70% U
T
(30% dip in U
T
)
for 25/30 cycles (for
0.5 sec)
0% U
T
(100% dip in U
T
)
for 250/300 cycles
(for 0.5 sec)
0% U
T
(100% dip in
U
T
) for 0.5/1 cycle
70% U
T
(30% dip in
U
T
) for 25/30 cycles
(for
0.5 sec)
0% U
T
(100% dip in
U
T
) for 250/300
cycles
(for 0.5 sec)
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the
Aoralscan 3
requires
continued operation during
power mains interruptions, it
is recommended that the
Aoralscan 3
be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
typical
commercial
or
hospital
environment.
If image distortion occurs, it
may be necessary to position
the
Aoralscan 3
further from
sources of power frequency
magnetic fields or to install
magnetic
shielding.
The
power frequency magnetic
field should be measured in
the
intended
installation
location to assure that it is
sufficiently low.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer
’
s declaration
–
electromagnetic immunity
Guidance and Manufacturer's Statement - Electromagnetic emission