Aoralscan 3_Manual_V1.0.0.30 - 19 -
150 kHz to 80 MHz
d
= 1.2
√P
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2.5 GHz
d
=
2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Table 2-5 Recommended separation distances between portable and mobile RF communications
equipment and the
Aoralscan 3
The medical electrical equipment is suitable for the professional healthcare environment per
60601-1-2:2014. It is suitable for use in physician offices, clinics, hospitals, and other professional
healthcare environments except near HF surgical equipment and the RF shielded room of an ME
system for magnetic resonance imaging or other environments where the intensity of
electromagnetic disturbances is high.
The clinical environments where the device can be used include physician offices, clinics,
hospitals, and clinical point-of-care for diagnosis of patients except environments where the
intensity of electromagnetic disturbances is high.
WARNINGS
⚫
Using accessories outside Shining3D provisions might lead to increased electromagnetic
emissions, or electromagnetic immunity reduced.
⚫
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Aoralscan 3, including cables specified by the manufacturer. Otherwise, degradation of the