Aoralscan 3_Manual_V1.0.0.30 - 43 -
8.
Care and Maintenance
Pre-cleaning, Disinfection, and Sterilization
The whole set of
Aoralscan 3
, including scanner head, scanner body, and scanner pedestal,
requires proper care, cleaning, and handling. As individual part may be processed differently, read
and follow the information and instructions given to help you effectively and thoroughly reprocess
the set.
We suggest that you reprocess the
Aoralscan 3
in the following order:
(1) Scanner pedestal care
(2) Scanner body care
(3) Scanner head care
WARNINGS
⚫
All parts are shipped non-sterilized. Follow the instructions prior to initial use.
⚫
Ensure that you have completely disconnected the power supply and all connections
from the scanner.
⚫
Follow the instructions given in the Manual to pre-clean, disinfect, and sterilize each
part of the scanner. Using other methods not approved by Shining3D Corporation will damage
your scanner and void your warranty.
⚫
Using detergent, disinfection solutions or wipes, sterilization procedures other than
those specified in the Manual may damage the product and void your warranty.
⚫
Only sterilize the part(s) for which a sterilization method is specified. Do not attempt to
sterilize all parts of the product. Shining3D Corporation is not liable for any damages due to
improper sterilization.
⚫
After sterilization, wait until each of the parts is at room temperature to prevent
possible heat injuries to the user and the patient.
⚫
To prevent cross-contamination, pre-cleaning, disinfection, and sterilization must be
correctly performed after each use.
⚫
When the scanner head is detached from the scanner, always protect the subtle units
and the inner optical components on the front end of the scanner body by putting on the supplied
protection cap.