41
1.
What if no scan data when the turntable has rotated one circle?
Solution: Please make sure that the object won’t block the mark points on the turntable. Or,
there will be no fringe pattern, while turntable will be still rotating. If the align mode is mark
point, please make sure that the marks on the turntable are covered, so as not to affect the
scanning. Make sure that in each single scan area has at least 4 points.
2.
What if the merging fails without mark points when the turntable has rotated several circles or
when it is under f ixed scan mode?
Solution: Try to make sure there are at least 1/3 overlap between the current scan area and the
previous scan area and the object surface should be featured. For objects which are symmetric
and without rich features, using mark points or manual merger is recommended.
3.
How to scan objects in transparent, semi-transparent or black?
Solution
:
Scan before spraying on the surface.
4.
Under auto scan mode, if the turntable is not moving, but with a humming sound, how to
solve?
Solution
:
Disconnect power line and connect again in few seconds.