16
4.2 Description of Functions
Item
Set Value
Description
Bar graph display
1. b.G.
0
Disable
☆
1
Enable
Auto-zero
(zero-tracking)
3. A.0
0
Disable
This function automatically sets the zero
point exactly to zero to prevent slight
deviations.
☆
1
Enable
Auto power-off
4. A.P.
0
Disable (balance operates
continuously)
This function is
available only when
the balance is
battery-operated.
☆
1
Enable (balance powers off in
approximately three minutes)
Response speed
5. rE.
0
Measurement by consecutive weighings.
1
Fast
Slow
2
☆
3
4
5
Stability parameters
6. S.d.
1
Wide (mild)
Narrow (strict)
☆
2
3
4
Interface
7. I.F.
0
Disable input/output
☆
1
Six-digit numeric format
2
Seven-digit numeric format
Setup of units of
measurement to be
displayed
Register selected
measuring units with
Function key.
81.S.u.
|
85.S.u.
☆
2 01 [ g ]
☆
1 02 [ ] (ct)
15 [ ] (oz)
16 [ ] (lb)
17 [ ] (ozt)
18 [ ] (dwt)
19 [
Lower right
] (grain)
1A
[ ] (tl_Hong Kong)
1b
[
Upper right] (tl_Singapore,Malaysia)
1C
[
Lower right] (tl_Taiwan)
1d
[mom]
1E
[ ] (to)
☆
3-5 00
Unit not set
Items marked
☆
are the default factory settings.
☆
1-
☆
5: default settings [81.S.u.]-[85.S.u.]
※
1
[00] cannot be set at [81.S.u.].