5 User information setting
FS series operation manual
-
44
-
5-9
Coefficient value setting
Can not be used.
The value that is obtained by multiplying a measured weight by a predetermined coefficient
can be indicated.
For example, in the case the coefficient is “2.35” and the measured weight is “2000 g”, a
value of “4700 g” is indicated.
(Example) Object to be weighed (2000 g)
×
coefficient (2.35)
→
Indication (4700)
1
Select a setting menu.
Select the coefficient scale.
(“2-7-2 Operation of the setting menu, setting of various
functions”)
2
Set the coefficient scale setting.
Push the [Direction] key.
Select “38.CF.”
Push the [Enter] key.
3
Set a coefficient value.
(“2-7-3 Operation of the setting menu, inputting of
numeric values”)
Input a coefficient value with the
[Numeric keypad].
Push the [Enter] key.
The coefficient value is stored.
4
The operation mode returns to the scale mode.
Returns to the scale mode with the
[Shift] and the [Clear Home] keys.
5
The object is weighed.
The value obtained by multiplying the
measured weight by the
predetermined coefficient is indicated.