6 External input / output functions
FS series operation manual
-
50
-
6-5
GZII format
This is different from “6-4 GZIII format” only in the operation of the T-command. In the GZII
format, the tare weight subtraction / zero adjustment is executed by the T-command. For
other specifications, please refer to “6-4 GZIII format”.
6-5-1
Command format
Please take care not to take alphabetical “O” for Arabic number “0.”
C1
C2
Code
(
C1
)
Code
(
C2
)
Description
Response
A00, Exx
format
A00, Exx
format
T
(SP) 0x54 0x20 Tare weight subtraction / Zero adjustment
A00
:
Normal
completion
E01
:
Command
error
E04
:
Tare weight
subtraction
/
Zero
adjustment
unavailable
ACK
:
Normal
response
NAK
:
Abnormal
response
O
0
0x4f
0x30 Output stop
A00
:
Normal
completion
E01
:
Command
error
O
1
0x4f
0x31 Continuous output at all times
O
2
0x4f
0x32 Continuous output at stable times
(Output stop at unstable times)
O
3
0x4f
0x33 Push down [Transfer] key for one-time
instant output.
O
4
0x4f
0x34 Auto output
O
5
0x4f
0x35 One-time output at stable times
(Output stop at unstable times)
O
6
0x4f
0x36 One-time output at stable times
(Continuous output at unstable times)
O
7
0x4f
0x37
Push down [Transfer] key for one-time
output at stable times.
O
8
0x4f
0x38
One-time instant output
O
9
0x4f
0x39 One-time output after stability is obtained