FS series operation manual
8 Controlling and adjustment functions
-
67
-
8-15
Reset to the factory settings
1
Select a setting menu.
Select the factory settings.
(“2-7-2 Operation of the setting menu, setting of
various functions” and “8-1 Hierarchy of controlling
and adjustment functions”)
2
Reset to the factory settings.
Inputting of the setting value
Push the [Direction] key.
Select “61E.IN.”
Input a setting value.
「
61E .IN. 0
」:
Not to be reset
「
61E .IN. 1
」:
To be reset
8-16
Span adjustment
Can not be used.
Span adjustment is to decrease the difference between an indicated value and the true value (mass).
This must be performed without fail in the case of doing high-accuracy weighing work.
Because an electronic scale is affected by the acceleration of gravity, adjustment is needed at every
weighing location. The adjustment is also needed when (1) using a long period and (2) an accurate
indication does not appear any longer.
(1) An external weight used for the span adjustment shall be the one equivalent to the
OIML F1 class.
(2) The span adjustment significantly affects the weighing accuracy. Please read this
procedure carefully before getting to the adjustment.
1
Select a setting menu.
Select the span adjustment.
(“2-7-2 Operation of the setting menu, setting of
various functions” and “8-1 Hierarchy of controlling
and adjustment functions”)
2
Select the span adjustment.
Push the [Direction] key.
Select “621.CE.”