3
Functions related to the operation
FZ-Ex series operetion manual
-
24
-
3-5
Comparator function
It is possible to preset threshold values and determine whether or not a measured value is
within the range defined by the preset values.
The comparator function can be used in any scale mode, i.e. weight scale mode,
percent scale mode, and coefficient scale mode.
3-5-1
How to perform discrimination
Set the lower and the upper limits. Then, whether or not the weight of an object to be weighed
is low (lower than the lower limit), appropriate or high (higher than the upper limit) is indicated
on the main LCD with
「
」
.
Single point
(lower limit) setting
Single point
(upper limit) setting
Two-point
(upper and lower limits) setting
Over the
upper limit
Appropriate
amount
Below the
lower limit
Over the
upper limit
Appropriate
amount
Below the
lower limit
Over the
upper limit
Appropriate
amount
Below the
lower limit
3-5-2
Discrimination criteria, and upper and lower limits setting
The discrimination is performed according to the following criteria:
Absolute value The discrimination is performed based on the upper and lower limit values that
have been set in advance.
Relative value
A reference numeric value is set in advance, and the discrimination is performed
based on the range defined by the upper and lower limit values that have been
set for the reference numeric value.
Summary of Contents for FZ100K1GEx
Page 2: ......
Page 101: ...FZ Ex series operation manual Appendix 91 Hierarchy of functions related to the lock ...
Page 106: ......
Page 107: ......
Page 108: ......