69
18 Use Printers
18.1 Setting up the Printer
(1) Use CSP-160 with the balance.
(2) Set proper print functions (print control) with the balance referring to the instruction manual
for your printer.
The factory default of our printer is manual printing (printer control).
(3) Make the baud rate and other settings compatible between the balance and the printer.
18.2 Outputting calibration results
1. Set
Function
1.
Set the ISO/GLP/GMP-compatible item to [F. GLP 1] and the GLP-compatible item for
measurement data to [F1. out 1].
2. Performs span adjustment/span test.
The data is output in a GLP-compatible manner
with [buSy
] displayed when span adjustment is
completed, or with the weight error remaining
displayed when a span test is completed.
While the data is being output, the balance may seem to not be moving but this does not
mean it has malfunctioned. Wait until the output is completed.
18.3 Output of Measurement Data in Compliance with ISO/GLP/GMP
1. Set
Function
1.
Set the ISO/GLP/GMP-compatible item to [F. GLP 1] and the GLP-compatible item for
measurement data to [F2. od. 1].
2. Header
output.
Press and hold the Print key for a few seconds.
[HEAd] is displayed and the header is then output.
3. Measurement data output.
Measurement data can be output at any time during measurements.
4. Footer
output.
Press and hold the Print key for a few seconds
after the measurement is completed.
The display indicates [Foot] and a footer is output.
Caution
(1) Upon printing calibration/span test results or ISO/GLP/GMP compatible outputs, date and
time data is also printed.
Check the date and time set in the balance before printing. (Refer to Section 15: Date
and Time Setup on page 49).
(2) Do not press any keys on the printer during printing. Refer to Section 19 “Output in
Compliance with ISO/GLP/GMP” described from page 70 for sample prints.
(3) While the data is being output in a special format, the measurement data cannot be
printed.
Continuous
press
Continuous