7 Troubleshooting
LF(R) Series Operation Manual
- 112 -
Message Cause
Remedy
1-Err
The reference weight used during
span adjustment or span test with an
external weight is far less than 50% of
the weighing capacity.
Use a weight that is as close to
the weighing capacity as
possible.
2-Err
An error over 1.0% was detected in
span adjustment or span test with an
external weight, or the balance failed.
Check that the correct weight
was put in place and that no
objects other than the weight are
on the pan. Then, execute span
adjustment or span test again.
b-Err
The balance is influenced by static
electricity or noise.
•
Unplug the AC adapter from
the receptacle and then turn
the power on again.
•
If this error occurs again,
electric components may have
failed. Please contact the
retailer from whom the balance
was purchased.
d-Err
The balance is influenced by static
electricity or noise.
•
Unplug the AC adapter from
the receptacle and then turn
the power on again.
•
If this error occurs again,
electric components may have
failed. Please contact the
retailer from whom the balance
was purchased.
L-Err
The weight of a sample is too light to
take samples for counting, or for
reference weight saving during
percentage weighing.
Use a heavier sample by
referring to the Specifications to
check the minimum unit weight
and the percentage weighing
weight limit.
t-Err
•
While using the addition operation,
two additional samples were placed
on the balance at once.
•
The addition operation was
performed when zero or a negative
value is displayed.
•
After returning the display to
“0” (by removing the previous
sample), place the next sample
on the balance to continue the
addition operation.
•
Addition operations are
impossible when 0 or a
negative value is displayed.
Place a sample on the balance
to continue the addition
operation.
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......