4 Adjusting the Scale
LF(R) Series Operation Manual
- 68 -
Setting “5A.rE.” to “1” results in the fastest response, which is suitable for use in a place
hardly affected by vibration. Setting “5A.rE.” to a larger number up to “7” results in lower
response, which is suitable for use in a place significantly affected by vibration.
Setting “5C.Fr.” to “0” results in the fastest response, which is suitable for use in a place
hardly affected by vibration. Setting “5C.Fr.” to a larger number up to “4” results in lower
response, which is suitable for use in a place significantly affected by vibration.
This function is not available when the addition function or the gravimeter mode is
used.
Summary of Contents for LF124R
Page 130: ...7 Troubleshooting LF R Series Operation Manual 120 Memo ...
Page 131: ...LF R Series Operation Manual Appendixes 121 Appendixes Appendix 1 Function Setting List ...
Page 132: ...Appendixes LF R Series Operation Manual 122 ...
Page 133: ...LF R Series Operation Manual Appendixes 123 ...
Page 134: ...Appendixes LF R Series Operation Manual 124 ...
Page 135: ...LF R Series Operation Manual Appendixes 125 ...
Page 136: ...Appendixes LF R Series Operation Manual 126 ...
Page 137: ...LF R Series Operation Manual Appendixes 127 ...
Page 138: ...Appendixes LF R Series Operation Manual 128 ...
Page 139: ...LF R Series Operation Manual Appendixes 129 ...
Page 144: ...Appendixes LF R Series Operation Manual 134 α check Display test α check Key test ...
Page 145: ...LF R Series Operation Manual Appendixes 135 α check Motor test Span adjustment history ...
Page 149: ...LF R Series Operation Manual Appendixes 139 Appendix 4 4 Unit Conversion Table ...
Page 150: ...Appendixes LF R Series Operation Manual 140 Appendix 4 5 Outline Drawings ...
Page 153: ......
Page 154: ......