11
5-4 Installing the weighing pan
Before use, attach the optional weighing pan or any weighing jig of 100g or heavier to the
installation boss.
Example of weighing pan installation
注 意
Note
If the mass of the weighing pan or weighing jig is heavier than the range of the zero
point adjustment to be performed at power-on, then tare is performed at power-on.
When tare is performed, the available weighing range narrows by the mass of the
tare, which means that "available weighing range = capacity - mass of the tare."
To prevent damage to the mechanical section, do not apply a torque of 0.5N-m or
higher to the clamping screws to attach to the weighing boss during tightening.
Also, do not push down on the clamping screws and bolts too strongly.
Installation boss
Clamping screws (M4x6)
Weighing pan
Summary of Contents for UD-1
Page 52: ...48 Appendix 3 Outlines Weight measuring sensor UF 620 UF 3200 ...
Page 53: ...49 Indicator ...
Page 54: ......
Page 55: ...はくし ...
Page 56: ...はくし ...