6-2 Zero point adjustment / tare
Zero point adjustment/tare
1
Press the Zero/Tare key during
weighing.
The zero point adjustment or tare is automatically performed depending on the load
applied to the weight measuring sensor. At this time, "->0<-" is lit while the zero point is
displayed.
Differences between zero point adjustment and tare
Zero point adjustment
Tare
Working load
Approx. 1.5% or less of the
capacity
Approx. 1.5% or more of the capacity
Measured value
after the operation
Gross weight
Net weight
Available weighing
range against the
capacity
Unchanged
Becomes narrower by the weight of the
tare
Available weighing range =
capacity (100% of the capacity)
Available weighing range = capacity -
weight of the tare
Symbol
None
"
(upper)"
"o-Err" is displayed when the total weight of the samples is more than the capacity.
When the samples are added after the tare, only the (net) weight of the added samples is
displayed.
In addition to the method of pressing the Zero/Tare key, zero point
adjustment and tare can also be executed by using the RS-232C tare
command (refer to "1. Zero adjustment/tare command" in Section 7-4-2) or
by using an external tare input from the connector (refer to Section 7-1
"Connector terminal No. and their corresponding functions").