28
7-4-2 Command formats
1 Zero adjustment/tare command
Command
body
Code
Description
Response
C1
C2
T
(SP)
54H
20H
•
Zero adjustment
•
Tare
A00: Normal termination
E01: Command error
E04: Unable to execute zero
adjustment/tare (out of scope,
weight error, etc.)
E09: Time-out error
2 Output control setting
Command
body
Code
Description
Response
C1
C2
O
*1
0
*2
4FH
30H
Output stop
A00: Normal termination
E01: Command error
O
*1
1
4FH
31H
Continuous output
•
After executing a command, the device retains the status until the next command is
input.
However, when the power source is turned off, the setting returns to function setting
values.
*1: Uppercase alphabetic character “O,” *2: Numeric “zero”
3 Span adjustment command
Command
body
Code
Description
Response
C1
C2
C
0
43H
30H
Cal key invalid
Span adjustment is
prohibited.
A00: Normal termination
E01: Command error
E02: Operation prohibited
E03: Interruption by operation
E04: Abnormal termination
C
3
43H
33H
Begin span adjustment.
•
It takes some time to respond since the device responds after it finishes
processing.
•
Once a C0 command has been sent, the device does not execute the C3
command even if a C3 command is subsequently sent. To cancel it, the function
should be displayed on the indicator or the power should be turned off.
•
The device does not execute a C3 command when the function on the indicator is
set to “5. CA. 0” (Cal key invalid).
•
The sensor does not send responses A01 or A02 for span adjustment commands
from the indicator.