41
10-2 Troubleshooting
Symptom
Cause
Action
Nothing appears on the
display even when the
balance is powered on.
The power supply is not
connected.
Check that the power supply is
connected.
If nothing is displayed even when
the power supply is properly
connected, then the electrical
components or the power supply
cable of the balance may have
failed.
Please contact the retailer from
whom the balance was
purchased.
The display flickers.
The weight measuring sensor may
be affected by an external
influence such as wind and
vibration.
Change the stability decision and
readability settings.
The displayed weight is
incorrect.
This error is caused by the balance
not being used for a long period of
time or being relocated.
Perform span adjustment.
The weight measuring sensors are
not horizontal.
Check the levelness before
installation.
The tare range is set or is not set.
Unload the sample from the
weighing pan and then zero the
readout to continue measurement.
An error still exists after
calibration.
The balance was affected by an
external influence such as wind
and vibration during the calibration.
There are possible effects from
vibration or wind. Take wind and
vibration-preventive measures or
change the installation location of
the balance and perform the
calibration again.
The M flashes
continuously flashes.
(for example, when the
[Zero/Tare] key is
pressed)
The balance may be affected by an
external influence such as wind
and vibration.
There are possible effects from
vibration or wind. Take appropriate
measures or c h a n g e t h e
i n s t a l l a t i o n l o c a t i o n o f t h e
b a l a n c e .
To reset to the initial
settings at purchase
It is possible to reset this product to
the initial settings. Refer to Section
10-3 "How to reset to the initial
state."