112
11.2 Standard Specifications
11.2.1 Measurement Range
An arbitrary setup is possible by the front keypad.
Digital indication accuracy rating is described in Table11.2.1 below under the Reference operating
conditions.
Reference operating conditions: Ambient temperature:
23 2
Ambient humidity:
55 10 %RH
Power supply voltage:
85 to 264 V AC
Power-supply frequency: 50/60 Hz 1%
Warm-up time:
30 minutes min.
(Performance is based on a state of no vibration or shocks.)
Terminal block reference junction compensation accuracy is not included in the digital indication
accuracy rating.
Terminal block reference junction compensation accuracy:
B, R, S, Au-Fe, PR40-20:
1
K, E, J, T, C, N, PL- , U, L:
0.5
Table 11.2.1 Digital Indication Accuracy Rating
Measurement
Indication (Digital)
Recording
(Analog)
Type
Range
Measurement range
Accuracy
Max.
resolution
Accuracy
-10 to 10 mV
(0.2% of rdg + 3 digits)
10
V
0 to 20 mV
(0.2% of rdg + 3 digits)
10
V
0 to 50 mV
(0.2% of rdg + 2 digits)
10
V
-200 to 200 mV
(0.2% of rdg + 3 digits)
100
V
-1 to 1 V
(0.1% of rdg + 3 digits)
1 mV
0 to 5 V
(0.2% of rdg + 2 digits)
1 mV
-10 to 10 V
(0.3% of rdg + 3 digits)
10 mV
D
C
vo
lta
ge
,c
ur
re
nt
In
p
ut
4 to 20 mA
(0.2% of rdg + 2 digits)
0.01 mA
0.0 to 1820.0
(0.15% of rdg + 1 )
400 to 600 ,
2
Accuracy is not guaranteed
within a range of 0 to less than
400
B
32.0 to 3308.0
(0.15% of rdg + 1.8 )
752 to 1112 ,
3.6
Accuracy is not guaranteed
within a range of 32 to less than
752
0.0 to 1760.0
(0.15% of rdg + 1 )
However, 0 to 100 ,
3.7
100 to 300 ,
1.5
T
he
rm
oc
o
up
le
R1
32.0 to 3200.0
(0.15% of rdg + 1.8 )
However, 32 to 212 ,
6.7
212 to 572 ,
2.7
0.1 /0.18
Digital indication
accuracy
(0.3% of
span)