9
VETROSON® Oxy-Gen™ System
External Components Description
Generator
Vent Ports -
Three rectangular vent ports are positioned in the front to accept
large particle filters.
Oxygen (O2) hose Outlet
-
This female fitting is at the lower left-hand side. The 3' hose from
the receiver tank connects to it.
On/Off Switch -
Controls power to machine
Hours Meter -
The hours meter tracks time while the unit is running.
It provides an indication as to when service intervals are due.
Hepa Air Inlet Filter Ports
-
Located on the back panel. Number varies by model. A Hepa filter
with a foam cap must be inserted into the round ports(s) on the
back panel of the unit.
Power Cord -
The power cord is designed for use on 120 vac/60 Hz electrical
systems and comes with a 3-pronged ground fault protected plug.
The 230V 60Hz (US) units are supplied with a 20 amp, 250V
slotted plug, Type #6-20P. This requires a 6-20R receptacle
For foreign shipments the plug will be removed and a plug of local
configuration should be installed. The green wire is the Earth
Ground and the black and white wires are for the AC Supply
Receiver Tank Components
Tank-
Holds approximately 30 liters of oxygen.
This allows the system to function without interruption when the flush is
used, when an ICU cage is being filled or when ventilators are in use (with
the 50 PSI models).
Hoses-
The 3’ hose carries oxygen from the generator to the receiver tank. The 9’
hose takes the oxygen from the receiving tank to the practice's central
manifold. The system is equipped with check valves to prevent back flow
from the manifold to the
VETROSON® Oxy-Gen™ System.
Tank Safe-Guard
Valve-
Allows continuous flow of oxygen when the
VETROSON®
Oxy-Gen™ System
is in operation
6-20R Receptacle – 230V, 60Hz (US)