background image

 10 

 

VETROSON® Oxy-Gen™ Generator 

Specifications 

 

Model 

Delivers 

Volts  Starting 

Amps 

Running 

Amps 

Power 

Watts 

Weight 

(Lbs.)* 

Width 

(Inches) 

Depth 

(Inches) 

Height 

(Inches) 

VGS2015 

20 PSI, 15 LPM 

120 

20 

920 

145 

25 

17 

38 

VGS5015 

50 PSI, 15 LPM 

230 

15 

1610 

175 

25 

17 

38 

 
 

*Weights shown are for the generator only. 
The Receiver Tank weighs an additional 34# 
The shipping weight will be more due to the box, packing material and pallet. 

 

 

 

Summary of Contents for VETROSON Oxy-Gen VGS2015

Page 1: ... Sheila Drive Tinton Falls NJ 07724 Phone 732 933 0800 Fax 732 933 0055 Toll Free 1 800 922 0722 sales summithilllaboratories com www summithilllaboratories com VETROSON Oxy Gen Systems Installation Operation Maintenance Manual ...

Page 2: ......

Page 3: ...INSTRUCTIONS 3 SHUTDOWN PROCEDURES 5 WARRANTY 6 LIMITS OF LIABILITY 7 PROBLEM SYMPTOMS 8 EXTERNAL COMPONENTS DESCRIPTION 9 GENERATOR 9 RECEIVER TANK COMPONENTS 9 VETROSON OXY GEN GENERATOR SPECIFICATIONS 10 ROUTINE MAINTENANCE INSTRUCTIONS 11 LONG TERM MAINTENANCE 11 TECHNICAL SERVICE AND ASSISTANCE 12 ...

Page 4: ......

Page 5: ...where it will reside Please note that one or more HEPA filters are packaged in bubbly in the oxygen generator box They should be firmly pressed into the port s at the back of the unit The smaller box will contain the receiver tank with the shorter 3 oxygen hose attached to the tank on one side The larger 9 oxygen hose will be attached to the tank on the other side These hoses will be wrapped aroun...

Page 6: ...bited near the unit Individuals who have experience handling oxygen systems should become the designated operators of the VETROSON Oxy Gen System within your facility 2 Ensure that the oxygen outlet stream is not directed toward anyone s clothing when bleeding a tank or line Oxygen could embed itself in the material and one spark or hot ash from a cigarette could ignite the clothing 3 It is import...

Page 7: ...l Laboratories Warranty Therefore if temperatures in the hospital or mobile hospital unit are outside of the peak performance zone the temperature of the hospital should be brought into the peak performance zone before operating the generating system ii The VETROSON Oxy Gen System must be operated in an upright position only and at least 12 inches 30 5 cm from any wall or obstruction to ensure tha...

Page 8: ...r 115V and 230V systems The 115V units are supplied with a 3 pronged ground fault protected plug The 230V 60Hz US units are supplied with a 20 amp 230V slotted plug Type 6 20P This requires a 6 20R receptacle Warning to owners of 230V Systems Many veterinary hospitals have a 209V triple phase electrical system This must be converted to 230V Single phase Our generator will produce oxygen using 209V...

Page 9: ...he LPM being utilized on the 2015 model After the initial set up the VETROSON Oxy Gen System will come up to specifications in about one minute after being turned on Note There is a circuit breaker button on the right side lower rear corner above the power cord It is an automatic shut off to protect the generator in case of power overload When operating your VETROSON Oxy Gen System please make sur...

Page 10: ...at the discretion of SHL excluding the molecular sieve elements will be affected The practice will pay shipping and insurance charges to Summit Hill Laboratories Summit Hill Laboratories will pay shipping charges back to the practice This Warranty shall also become null void and not binding on SHL if a defect or malfunction occurs in the product or any part of the product as a result of user error...

Page 11: ...nally a failure to clean and replace the filters will restrict air flow to the generator and may also cause a breakdown of the molecular sieve Under no circumstances is the molecular sieve covered under warranty by SHL Transfer This Warranty may be transferred to subsequent owners of the VETROSON Oxy Gen System only with the prior written approval of SHL The warranties above are given expressly in...

Page 12: ...he filters may be clogged The sieve beds may be contaminated due to excess humidity ie water or oil introduced introduced into the atmosphere Also on an older machine this may be an indication that the compressor has expended its useful life The ball on the external flowmeter is swinging up and down This may be an indication that 1 The practice is asking the unit to supply more LPM than it is desi...

Page 13: ...lt protected plug The 230V 60Hz US units are supplied with a 20 amp 250V slotted plug Type 6 20P This requires a 6 20R receptacle For foreign shipments the plug will be removed and a plug of local configuration should be installed The green wire is the Earth Ground and the black and white wires are for the AC Supply Receiver Tank Components Tank Holds approximately 30 liters of oxygen This allows ...

Page 14: ...Weight Lbs Width Inches Depth Inches Height Inches VGS2015 20 PSI 15 LPM 120 20 8 920 145 25 17 38 VGS5015 50 PSI 15 LPM 230 15 7 1610 175 25 17 38 Weights shown are for the generator only The Receiver Tank weighs an additional 34 The shipping weight will be more due to the box packing material and pallet ...

Page 15: ... The cabinet and power cord should occasionally be washed down with a clean sponge or clean rag and some soapy water Do not use ammonia or other strong chemical based cleaning solvents Avoid a buildup of dust and dirt on the machine The inside of the cabinet should also be inspected weekly and vacuumed if dust is noted Long Term Maintenance Air Compressor Air compressors are an important part of t...

Page 16: ...X or Common Carrier at Summit Hill Laboratories 3 Sheila Drive Tinton Falls NJ 07724 USA Shipping Units for Repairs A sturdy box must be used We recommend keeping the original shipping box The box must be strapped on a pallet in an upright position Bubbly packing material is mandatory Peanuts shredded paper and such loose and dusty material should not be used as they will migrate into the working ...

Reviews: