ADM-880C 07/20/09
6
3.3 FEATURES ON SIDES AND BACK OF METER
BATTERY CHARGER JACK
When viewed from the front, the battery charger jack is on the right side of the meter toward the top. The battery
charger plug is to be connected here.
EXTERNAL READ JACK
When viewed from the front, the external read jack is on the left side of the meter toward the top. The plug for the
external thumbswitch is connected here. This feature allows the operator to trigger measurements from the
FlowHood or VelGrid handgrip while working overhead or in awkward circumstances. The thumbswitch performs
the same function as the READ key.
FLAPS JACK
The flaps jack is on the back of the meter, in the upper right hand corner. The flaps plug on the FlowHood is inserted
here.
TEMPERATURE INPUT JACK
The temperature input jack is centered on the back of the meter, slightly toward the top. The flexible TemProbe
sensor must be connected to this receptacle whenever temperature density correction is desired for either flow or
velocity measurements. A retractile cord connects the TemProbe or the MultiTemp to the temperature input jack
for remote temperature sensing.
RESET SWITCH
The reset pushbutton switch is on the back of the meter in a recess near the upper left corner. This switch is used
to reset the meter in the unlikely event that the microprocessor becomes lost in its program. This may occur if the
meter is dropped, and may cause the keypad to become nonfunctional until the meter is reset. Press the reset switch
for one second to restart the meter.
Readings saved in memory since the meter was last turned on may be lost if the meter must be reset. If the meter
continues to fall into "lockout" after being reset, it may have been damaged, and should be returned for repair.
PNEUMATIC PRESSURE INLETS
Two pneumatic pressure inlets positive (+) and negative (-) are centered on the back of the meter at the top edge
and may be connected to various pressure sources for the measurement of air velocity, flow, or pressure. Sources
include the FlowHood, AirFoil probe, VelGrid, pitot tubes, static pressure probes, or any other pressure source not
exceeding the safe limits for the meter. The negative (-) inlet senses the static pressure during flow or velocity
measurements, and also is used for direct absolute pressure measurements.
SERIAL PORT JACK
When viewed from the front, the serial port jack is centered on the right side of the meter just below the battery
charger jack. The circular plug of a custom RS232 serial cable is connected here. The other end of the cable
provides a standard DB9 connection for a printer or computer.
Summary of Contents for AIRDATA ADM-880C
Page 42: ...ADM 880C 07 20 09 37 FIGURE 6 3 VELGRID ASSEMBLY...
Page 50: ...ADM 880C 07 20 09 45 FIGURE 10 1 FRAME STORAGE FIGURE 10 2 FLOWHOOD IN CASE...
Page 51: ...ADM 880C 07 20 09 46 FIGURE 10 3 FLOWHOOD ASSEMBLY...
Page 53: ...ADM 880C 07 20 09 48 FIGURE 10 7 1X5 FRAME ASSEMBLY FIGURE 10 8 3X3 FRAME ASSEMBLY...