37
8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
ASSEMBLY AND INSTALLATION
3
Fig. 19: Aligning the flanges with tube
▸
Finally, weld the flange tubes tight all-round and, at the same time, check for exact
alignment and correct when necessary. When using the adjusting device, assemble??
both the flange plate with light source and the flange plate with target optics before
welding the second flange tube.
▸
Close off the flange opening after fitting to prevent gas escaping.
Aligning with adjusting device
(
see “Options for MCU control unit”, page 123
; also available on loan)
A
A
Aligning with auxiliary tube
Auxiliary tube
Duct wall
Flange with tube
Light source
Use the target optics to align the flange so that the light spot of the lamp appears in
the center of the target optics.