8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
4.2.2
Focussing the sender light beam for transmission measurement
▸
Select the “Adjustment / Manual Adjustment / Transmission set reference” directory
and activate the “Permanent LED light” checkbox in the “Adjustment aids” field.
Fig. 36: SOPAS ET menu: DH C200/Adjustment/Manual Adjustment/Transmission set reference
▸
Loosen the knurled screws (
see “Sender/receiver unit DHC-T for transmission and scat-
tered light measurement”, page 18
), swivel the electronics unit to the side and remove
pivoted shutter cover (1).
Fig. 37: Cover screw for focussing opening