8011952/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER C200
Subject to change without notice
4
START-UP AND PARAMETER SETTINGS
4.3
Sender/receiver unit and reflector/scattered light receiver
After completion of the tasks described above, the sender/receiver unit and the reflector/
scattered light receiver must be removed from the adjusting stands or zero tube and taken
to the measuring location.
4.3.1
Connecting the scattered light receiver
▸
Check whether the purge air supply is available (the flow direction must be correct and
the purge air hoses fitted tight on the connections).
▸
For purge air supply via the MCU-P control unit or external purge air supply, push the
purge air hose on the connections of the sender/receiver unit and the reflector and
secure with hose clamps.
4.3.2
Fitting and connecting on the duct
▸
Attach the sender/receiver unit and reflector/scattered light receiver to the flange with
tube and secure with the associated assembly kit (
see “Assembly parts”, page 122
tighten the self-locking nuts as tight as possible.
Fig. 42: Fitting the sender/receiver unit / reflector on the duct
C
A
B
1 pair
Sender/receiver unit
Self-locking nut
Spherical washer
Sealing tape
Cup springs (4 pairs); only with fastening set for sender/receiver unit
A
Horizontal alignment
B
Fixing point
C
Vertical alignment