Maintenance
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
163
14.4.2
Replacement/cleaning of the filter element (procedure)
Activate the “Filter change” status:
1
If the display shows the measurement display:
Press
to open the main menu.
2
Open the “Filter change” menu: Main menu
Maintenance
Filter change.
3
If the “System in Service” status has not yet been activated:
Activate “System in Service” in the displayed menu (
➝
The “Filter change” menu is displayed:
4
Press
or
.
➝
Selection box 1 is highlighted.
5
Press .
➝
The selection switches to
Y E S
.
➝
Selection box 2 is highlighted.
6
Press
or
.
➝
E n d
is highlighted.
7
Press .
➝
The measuring system is protected against soiling (for explanation refer to
Maintenance work is easier and faster if you have a completely cleaned filter
element for replacement (
).
Measuring operation is interrupted during maintenance.
Inform any affected stations.
If necessary for safety purposes:
Deactivate connected devices.
NOTICE:
Only open the sample gas filter when the “Filter change” status is activated.
Otherwise dust-laden ambient air can enter the measuring system. Soiling
which occurs as a result can render the measuring system useless.
M a i n t e n a n c e
F i l t e r c h a n g e
F i l t e r c h a n g e
N O
F i l t e r i s r e p l a c e d
N O
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
S y s t e m i n S e r v i c e
Menu title
Selection box 1
Selection box 2
Current measured value
Current measuring range (measuring range end value)
M a i n t e n a n c e
F i l t e r c h a n g e
F i l t e r c h a n g e
Y E S
F i l t e r i s r e p l a c e d
C h a n g e o r E n d
6 1 . 6 m g C
R a n g e . . . 1 0 0
S y s t e m i n S e r v i c e
Menu title
Selection box 1
Selection box 2
Current measured value
Current measuring range (measuring range end value)
Back
E n d
N O *
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...