172
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Clearing Malfunctions
15.3.3
Fuses in the terminal box
Fig. 30
Fuses in the terminal box
Identifier
Mains voltage
Fuse
Order No.
F100
115V
Fuse link T0A8 250V D5x20
6004292
230V
Fuse link T0A4 250V D5x20
6004278
F101
F102
115V
Fuse link T4A0 250V D5x20
6004310
230V
Fuse link T2A0 250V D5x20
6004303
Older versions may have other values.
Comply first with the labeling on the terminal box.
Older version of the terminal box
F100
F101
F102