EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Analyzer Unit
4.2
Installation of the “19" slide-in enclosure” type analyzer unit
4.2.1
Installation/securing
Ambient conditions
Do not block the fan openings.
Ensure a supply of cool air (ambient air, fresh air).
Comply with the permissible ambient temperature (
).
Assembly
Install the operating unit type “19" slide-in enclosure” in the normal manner in a 19" frame
or a corresponding outer housing
Fig. 5
Analyzer unit type“19" slide-in enclosure”: Recommended installation position
The notes in this section are based on the assumption that the analyzer unit
type “19" slide-in enclosure” is combined with an operating unit type “19"
slide-in enclosure” with an integrated terminal box.
NOTICE:
Do not only secure the housing by the front panel.
Use the rack rails which bear the weight of the device.
Otherwise the device would be damaged.
Do not install the operating unit directly above the analyzer unit (
).
This has a favorable effect on the temperature conditions in the operating
unit.
Mount the 19" slide-in enclosures in a pivoting frame. This facilitates
access to the electrical connections and mains fuses.
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...