Installation of the Analyzer Unit
4.3
Installation of the “inline” type analyzer unit
4.3.1
Safety notes on the lifting points
Always handle and store the analyzer and the complete analyzer unit so that no forces
act on the sampling probe.
Only lift the analyzer unit by the base frame, not by the sampling probe (
).
Fig. 7
Lifting points on the EuroFID analyzer
NOTICE:
The surrounding tube of the sampling probe contains a glass vacuum tube.
Do not subject the sampling probe to mechanical loads.
Do not use the sampling probe as a lifting point.
Otherwise the vacuum tube can break.
If the vacuum tube breaks, condensate can form in the sampling probe during
operation. Condensate in the sample gas path can damage the measuring
system or render it useless.