Installation of the Analyzer Unit
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
49
4.4.2
Assembly
Installation site
●
Ambient temperature:
– Standard version (with fan): –30 … +30 °C
– Version with air conditioner (option): –30 … +50 °C
●
The installation area must be free from vibrations and jolts.
●
The sampling probe protrude freely into the process gas flow.
Preparing for assembly
Install a suitable mounting connection at the planned installation site to which the
analyzer unit can be attached.
Take the space requirements of the housing into account (dimensions
).
Check whether the flange requires support to bear the weight of the device. If
necessary support the flange.
4.4.3
Installation of the industrial housing
Secure the industrial housing to the mounting connection (gas connections pointing
downwards).
Fit a flange seal between the industrial housing and the mounting connection.
NOTICE:
The bypass exhaust gas flow is discharged from the surrounding tube of the
sampling probe.
For detailed information refer to
The analyzer unit type “industrial housing” contains a frost-protection heater.
●
Flange specification
●
Suitable DIN mounting connection (accessory)
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...