92
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Main Menu Functions
10.4
Flame ignition
Function
This menu function is used to manually ignite the FID flame. This is only necessary in the
event of maintenance work or for troubleshooting.
When ignition is started, the fuel gas supply is briefly increased and the FID flame is ignited
with a spark plug. If this does not work immediately, the EuroFID automatically makes
several ignition attempts. If 10 consecutive ignition attempts were unsuccessful, a
malfunction message is displayed.
When you start the function while the FID flame is burning, nothing happens.
Procedure
1
Open the menu: Main menu
Flame Ignition
To start the function:
2
Press .
➝
The main menu appears – with the added information
I g n i t i o n
.
F l a m e I g n i t i o n
F l a m e I g n i t i o n
Y E S
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Selection box
Current measured value
Current measuring range (measuring range end value)
Status message
Status message
To terminate the function:
2
Press
or .
➝
The selection box is highlighted.
3
Press .
➝
The selection box switches to
N O
.
4
Press
or .
➝
E n d
is highlighted.
5
Press .
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...