“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
109
11.5.2
Displaying the error archive
Function
The “Error archive” menu is used to display the saved error messages on the display. Two
messages are displayed on each “page”.
Procedure
1
Open the menu: Main menu
Maintenance
Error archive
Disp. of err. archiv.
➝
Existing error messages are displayed in the sequence of their occurrence (the
latest first); two error messages are displayed per display page.
Example:.
➝
If additional error messages are provided,
< C o n t i n u e >
is displayed.
2
< C o n t i n u e >
or
< E N D >
: Press
.
To cancel: Press
.
●
Printing all error messages
●
Deleting all error messages
●
Error messages in full detail (diagnosis function)
D i s p . o f e r r . a r c h i v
1 5 . 0 5 . 2 0 0 3 0 9 : 5 2 : 3 1
M a i n t e n a n c e 0 0 0 7 9
S e r v i c e r e q u i r ( D i a g )
1 5 . 0 5 . 2 0 0 3 0 9 : 5 2 : 3 3
F i l t e r e r r o r 0 0 1 0 7
S e r v i c e r e q u i r ( D i a g )
1 . 2 p p m
R a n g e . . . 1 0 0
Menu title
Date/Time
Message / localization
1
Automatic reaction
Date/Time
Message / localization
Automatic reaction
Current measured value
Current measuring range (measuring range end value)
1
Number of the program section in which the error was localized
< E N D >
Back
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...