“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
115
11.9
Range select mode (Measuring range selection)
Function
In the “Range select mode” you can stipulate the required type of measuring range switch-
over. It is also possible to permanently program a certain measuring range.
Open the menu
Open the menu: Main menu
Maintenance
System Parameters
User level
Range
select mode.
Permanent setting of a specific measuring range
1
Press
or
until the output field is highlighted.
2
Press
until
F i x . r a n
is displayed.
3
Press
or
until the entry field is highlighted.
4
Press the numeric key which corresponds to the required measuring range (1 … 6).
Example:
To set measuring range 2, press [2].
To finish:
5
Press
or
until
E n d
is highlighted.
6
Press .
Table 11
Measuring range switch-over modes
F i x . r a n
Permanent setting of a measuring range
1
1
“Permanent setting of a specific measuring range” (P. 115)
M a n u a l
Manual measuring range selection (
a u t o
Automatic measuring range switch-over
2
; the analog output
“Reference value” shows measured values (
2
“Settings for automatic measuring range switch-over” (P. 116)
a u t o A O u
Automatic measuring range switch-over
; the analog output
“Reference value” signals the current measuring range (
e x t e r n a l
without function
R a n g e S e l e c t M o d e
M e a s u r i n g r a n g e s w .
C h a n g e o r
F a c t o r y l e v e l a c t i v e
R a n g e . . . 1 0 0
Menu title
Selection box (
)
Current measured value
R a n g e S e l e c t M o d e
M e a s u r i n g r a n g e s w .
M e a s u r i n g r a n g e n o
1 . . . 6 2
C h a n g e o r E n d
F a c t o r y l e v e l a c t i v e
R a n g e . . . 1 0 0
Menu title
Selection box (selectable modes
Entry field
End value of the set measuring range
Current measured value
Mode
Setting functions depending on
the mode
E n d
f i x . r a n
1 0 0 . 0 0 p p m
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...