11.11.1
Integrated value (averaging for the measurement display)
Function
The “Integrated value” is the interval of sliding averaging for the measurement display. This
produces a more constant measurement display if the measurement signal is irregular or
“noisy”. The EuroFID then does not display current measured values, but mean values from
the current measured value and the previous measured values.
Measured value damping is not active during calibration.
On devices which are specified as gas detectors, the integrated value is fixed at “0”
(cannot be adjusted).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
I/O-
Parameters
Integrated value
2
Press
or .
➝
Entry field is highlighted.
3
Enter the required value.
4
Press
or .
➝
E n d
is highlighted.
5
Press .
●
If the “Integrated value” is increased, the reaction time (90% time) of the
measurement display is greater.
●
If the “Integrated value” is decreased, the measurement irregularity of the
measurement display can increase.
There is also a similar setting option
– for the “Measured value” analog output
– for the “Reference value” analog output
Small particles can cause extreme fluctuations in the measured value. The
installed sample gas filter cannot retain particles which are smaller than 1
μ
m.
The particles burn spontaneously in the FID flame and cause fluctuations in
the measured value.
S y s t e m P a r a m e t e r s
I n t e g r a t e d v a l u e
F i x e d P
M e a s u r i n g r a n g e
1 . 0 s
u p : 0 . 0 s
u p t o : 6 0 . 0 s
I n t e g r a t i o n t i m e
P r e s s u r e
5 s
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Entry field
Minimum value
Maximum value
Averaging interval for integrated pressure sensors
1
Current measured value
Current measuring range (measuring range end value)
1
Do not change the works setting
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...