“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
127
11.11.10
Setting the measuring ranges
Function
The measuring ranges of the EuroFID can be changed by adjusting the end value of
measuring range 6 (setting range: 0.01 … 9999999.00). The end values of the smaller
measuring ranges are automatically modified (in the ratio 1:10).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
Scale
fixing.
2
Press
or
until the entry field is highlighted.
3
Enter the required measuring range end value.
To finish:
4
Press
or
, until
E n d
is highlighted.
5
Press .
●
The end values of the measuring ranges 1 to 5 cannot be individually
adjusted.
●
The beginning values of the measuring ranges are always “0” (not variable).
●
The physical unit cannot be changed in this menu.
NOTICE:
If the measuring range setting was changed:
Recalibrate the device.
S c a l e f i x i n g
M a x 2 3 7 4 4 p p m
U s e . s k a
u p u p t o
p p m p p m
0 . 0 0 . 0 7 5
0 . 0 0 . 7 5 0
0 . 0 7 . 5 0 0
0 . 0 7 5 . 0 0 0
0 . 0 7 5 0 . 0 0 0
0 . 0
C h a n g e o r E n d
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Maximum displayed measured value
1
Beginning value/end value of measuring range 1
… of measuring range 2
… of measuring range 3
… of measuring range 4
… of measuring range 5
Entry field: End value of measuring range 6
Current measured value
Current measuring range (measuring range end value)
1
If the measured value exceeds this value, “OFL” (overflow) is displayed as a measured value
M e a s r a n g e = 6
7 5 0 0 . 0 0 0
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...