148
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Calibration
12.4
Settings for calibration
12.4.1
Calibration control
Function
In the “Calibration control” menu you can determine the start and sequence of calibration
procedures and also decide whether the “Calibration block” control input should serve to
start automatic calibration.
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
Calibration control
Calibration control
S y s t e m P a r a m e t e r s
C a l i b r a t i o n c o n t r o l
C a l . M o d
m a n u a l
M R Z e r o
1
M R S p a n
2
M a x c a l i b d u r a t i o n
1 - 3 6 0 0 1 2 0 0 s
A u t o . C a l . b e f . m e a s u r
< N O >
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Selection box 1
Entry field 1
Entry field 2
Entry field 3
Selection box 2
Current measured value
Current measuring range (measuring range end value)
Table 15
Explanations for the “Calibration control” menu
C a l . m o d
M a n u a l
Calibration is only started and monitored by the
operator with a menu function
1
1
= Standard setting set at manufacturer's works
a u t o
Automatic calibration starts automatically at the set
times (
) (manual calibration is still
possible)
e x . a u t o
Automatic calibration is started via the “Calibration
block” control input (
e x . S t e u
without function
M R Z e r o
Measuring range in which the zero gas is measured
2
2
Recommendation: Smallest measuring range used in measurement mode. Instrument air must be pure enough for
the selected measuring range (specifications
M R S p a n
Measuring range in which the span gas is measured
3
3
“Typical setting” for emission measurement: 2: for LEL measurement; 5
M a x c a l i b
d u r a t i o n
Specifies how long the EuroFID tries to find a constant measured
value (average value) with a calibration gas which can be used for
calibration. If this is not possible within the “Max. calib duration”,
calibration is canceled and a malfunction message is activated.
A u t o . C a l .
b e f o r
m e a s u r
Y e s
After each start-up (also after stand-by), the EuroFID
starts automatic calibration as soon as the
operational state is reached.
N O
No automatic calibration starts after start-up.
4
4
Recommended setting / for emission measurement: Obligatory setting
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...