The gas analyzer EuroFID is used to measure the total carbon content of hydrocarbon
compounds in air.
●
The possible measured concentration ranges from 0.015 ppm to 100 % LEL.
●
The EuroFID can analyze corrosive and condensing gases at gas temperatures up to
350 °C (660 °F).
●
The EuroFID is tested for suitability for continuous monitoring of the emissions of
organically bound carbon according to
– 13. FICA and TI Air
– 17. FICA
1.4.2
Installation site
●
The operating unit and analyzer unit (
) are generally designed for indoor
operation. These devices should not be exposed to direct atmospheric influences (wind,
rain, sun). Such influences can damage the devices
and negatively affect the
measuring accuracy.
●
The “inline” type analyzer unit can also be used outdoors if it is provided with a hood for
weather protection (accessory).
●
The “industrial housing” type analyzer units can be used outdoors.
●
The terminal box can be used outdoors.
●
The EuroFID is not suitable for operation in potentially explosive atmospheres.
The maximum permissible distance between the terminal box and the analyzer
unit is 500 m. The standard cable length is 6 m.
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...