Clearing Malfunctions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
171
15.3
Electrical fuses
15.3.1
Mains fuse of the operating unit
Fig. 29
Mains fuse of the operating unit
15.3.2
Fuses in the analyzer unit type “industrial housing”
The analyzer unit type “industrial housing” has two additional mains fuses integrated into
terminals 8 and 9 of the connection terminal strip (
CAUTION:
Health risk
If the housing of the mains fuses is removed, electrical contacts are exposed
which carry mains voltage.
Before testing the fuses:
Disconnect the EuroFID from the mains supply or
switch the power supply off at an external point.
CAUTION:
Fire hazard / danger of irreparable damage
If incorrect fuses are fitted, a fire can occur in the event of a defect.
Only use replacement fuses which correspond exactly to the stated ratings
(design, disconnecting current, disconnecting characteristic).
Identifier
Fuse
Order No.
Fuses
Fuse link M4A0 250V D5x20
6000471
Power
Netz
Fuses
Sicherungen
4A MT
Identifier
Fuse
Order No.
F1
Fuse link T4A0 250V D5x20
6004310
F2
Fuse link M1A0 250V D5x20
6004296
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...