Product Description
2.3
Measuring ranges
Measuring ranges
The EuroFID has 6 measuring ranges. Each measuring range starts at “0”. The measuring
range end values vary by the factor 10. The physical unit can be selected (
).
The measuring ranges can be modified by changing the end value of the largest measuring
range (
). The smaller measuring ranges are automatically derived from
the largest measuring range and are therefore not variable.
Alternative scaling
The EuroFID is calibrated at the works with propane, i.e. the hydrocarbon measurement is
modified for propane.
It is possible to reference the measuring ranges to another physical unit (e.g. mg/m3) or
another substance (e.g. butane). To this purpose corresponding conversion factors can be
set (
).
Table 1
Measuring ranges (example)
Measuring range 1
0 … 2 ppm
Propane
Measuring range 2
0 … 20 ppm
Measuring range 3
0 … 200 ppm
Measuring range 4
0 … 2000 ppm
Measuring range 5
0 … 20000 ppm
Measuring range 6
0 … 200000 ppm
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...