EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Project Planning
3.2.2
Mechanical installations (overview)
The following system components must be installed:
3.2.3
Mains connection (overview)
●
The mains voltage for all device modules is supplied through the operating unit (
).
●
An additional, separate mains switch must normally be installed (
).
Component
Installation site, requirements
Detailed notes
Supply line or
compressed air
cylinders for
instrument air
– Suitable pressure reducer
– Gas line to the analyzer unit
– Oil-free installation
Compressed gas
cylinder(s) for fuel
gas
– At a suitable location which meets the
safety requirements
– Suitable pressure reducer
– Gas line to the analyzer unit
– Leakage monitoring device
1
1
(recommendation)
Analyzer unit
–
“Wall housing” type:
In an enclosed
interior (e.g. system cabinet)
–
“19" slide-in enclosure” type:
In 19"
frame
–
“Inline” type:
On a mounting connection
–
“Industrial housing” type:
On a mounting
connection
Shut-off valves
– Near the analyzer unit
2
2
Pre-assembled shut-off fitting available as accessory
Refer to “Analyzer
unit”
Solenoid valve
3
3
Only for devices with the “flow barrier” option
– Near the analyzer unit
Terminal box
4
4
Not for device versions with an integrated terminal box
– Near the analyzer unit
5
5
Standard length of the connection cable: 3 m
Operating unit
– Near the
6
external terminal box
or the
analyzer unit
6
Standard length of the connection cable: 6 m
Electrical
connections
– Analyzer unit (
terminal box)
Analyzer unit
– Signal connections as required
The terminal box receives the mains voltage from the operating unit and in turn
supplies the analyzer unit with operating voltage.
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...