54
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Terminal Box
5.1
Installing the terminal box
Mount the terminal box near the analyzer unit (take the cable length into account).
Select the installation position so that the cable ducts are on the underside of the
terminal box.
5.2
Electrical connections of the terminal box
Connection cable of the analyzer unit
Connect the connection cable of the analyzer unit to the respective plug connector of
the terminal box.
Connection cable to the operating unit
Does not apply if the terminal box is integrated into the operating unit.
Connect both connection cables of the terminal box to the operating unit (plug
connectors “Analyzer assembly” and “Data cable in”).
Customized connection cables
For customized connection cables use the following cable material:
Establish the wiring connections according to
(
Connection for flow barrier
Only applies to devices with the “flow barrier” option.
Does not apply if the terminal box is integrated into the analyzer unit.
Connect the solenoid valve of the flow barrier to the connections “VALVE (+24 V)” (X19)
and “0 V” (X18).
Connection for housing fan
Only applies for devices with the “fan” option.
Connect the connection cable of the housing fan to a “+24 V”/“0 V” connection (X21/
X20 or X37/X36).
CAUTION:
Damage/malfunction caused by incorrect mains voltage
The terminal box is supplied with mains voltage directly from the operating
unit. If the supplied mains voltage does not comply with the mains voltage
configuration of the terminal box, the terminal box and the EuroFID can be
damaged. If damaged, the devices can constitute a hazard.
Only run the terminal box with mains voltage which complies with the
information on the type plate.
Connection
Wire cross-section
Type
Recommended material
Analyzer assembly
terminal box
1
1
carries mains voltage
3 x 1.5 mm
2
shielded
Ölflex Classic 110 CY
Data cable in
data line
4P x 0.34 mm
2
AWG22
UL style no. 2464/1061
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...