58
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Operating Unit
6.1
Installing the operating unit
Installation site
Do not obstruct the fan openings (rear of unit).
Ensure a supply of cool air (ambient air, fresh air).
Comply with the permissible ambient temperature (
).
Installation of the operating unit type “19" slide-in enclosure”
Install the operating unit type “19" slide-in enclosure” in the normal manner in a 19"
frame or a corresponding outer housing
Use the rack rails.
Installation of the operating unit type “½-19" slide-in enclosure”
Install the operating unit type “½-19" slide-in enclosure” into a 19" housing frame
(4 HU).
Fig. 17
Operating unit: Installation of the operating unit type “½-19" slide-in enclosure”
Housing dimensions and space requirements:
– Operating unit type “19" slide-in enclosure”
– Operating unit type “½-19" slide-in enclosure”
NOTICE:
Do not secure the housing by the front panel alone.
Use the rails which support the housing.
Otherwise the device would probably be damaged.
Recommendation:
Mount the 19" slide-in enclosures in a pivoting frame. This facilitates
access to the connections and mains fuses.
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...