FLOWSIC100 Flare-XT · Operating Instructions · 8023761/V1-0/2020-10 · © SICK Engineering GmbH
Technical data
12.1
System FLOWSIC100 Flare-XT
NOTICE:
The exact device specifications and performance data of the product may
deviate and depend on the respective application and customer specification.
Only the metrological parameters described in the Application Evaluation
Datasheet apply.
If the delivery documentation of your FLOWSIC100 Flare-XT does not include an
Application Evaluation Datasheet, contact your SICK partner!
Example of an Application Evaluation Datasheet:
Table 22
System FLOWSIC100 Flare-XT
Measuring parameters
Measured values
Mass flow, volumetric flow s. c. (standard condition), volumetric flow a. c. (actual con
-
dition), molecular weight, gas volume and mass, gas velocity, gas temperature, sound
velocity
Number of measuring paths
1 path, 2 paths
Nominal pipe size
1-path measurement: 4 ″ ... 72 ″
2-path measurement: 12 ″ ... 72 ″
* Other nominal sizes on request
Measuring principle
Ultrasonic transit time difference measurement, ASC technology
Measured medium
Typical flare gas
Measuring ranges
1)
0.03 m/s … 120 m/s
Repeatability
According to ISO 5725-1; JCGM 200:2012): < 0.5% related to the measured value in
the range ≥ 1 m/s
Resolution
(According to JCGM 200:2012): + 0.001 m/s
Uncertainty of measurement
1,
2)
,
3)
Volumetric flow a. c.
1% … 5%
Related to the measured value with ultrasonic technology
(in the range ≥ 0.3 m/s to measuring range end value)
0.5% … 1.5% with spool piece and flow calibration
Related to the measured value with ultrasonic technology
(in the range ≥ 1 m/s to measuring range end value)
4)
Mass flow rate
2% … 5.5%
Related to the measured value with ultrasonic technology
(in the range ≥ 0.3 m/s to measuring range end value)
1.5% … 2% with spool piece and flow calibration
Related to the measured value with ultrasonic technology
(in the range ≥ 1 m/s to calibration range end value)
4
Uncertainty of measurement ASC
technology
1),2),
5)
Volumetric flow a. c.: 1% … 8%
Resolution
+ 0.001 m/s
Ambient humidity
≤ 95% relative humidity
Compliances
ATEX: 2014/34/EU
EMC: 2014/30/EU
RoHS: 2011/65/EU
PED: 2014/68/EU
Electrical safety
IEC 61010-1 (Non-Ex Interface Unit)
1)
Depending on application conditions such as gas composition, process temperature, device type, pipe diameter, etc.
For mass flow rate, additional selection and configuration of the conversion algorithm as well as uncertainty of the pressure and temperature
sensors. To be evaluated by SICK.
2)
With fully developed turbulent flow profile. Typically 20D straight inlet and 5D straight downstream piping is required .
3)
Below a specific threshold Reynolds number, only run time effects and uncertainties of geometry, excluding contributions from the flow profile
are considered.
4)
Depending on the capabilities of the selected flow lab.
5)
Additional uncertainty. In the range 100% ... 130% of the last gas velocity measurable with ultrasonic transit time difference measurement.