Maintenance
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
175
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Specify the "Collection duration", a timespan, over which live meter data is to be
collected to document the meter‘s state (default: 1 minute).
Enter the current pressure, temperature and SOS. Use the SOS Calculator to calculate
the SOS for the gas composition (
). The gas composition must be current
and representative.
Click the "Start" button to start live data collection. Diagnosis data, measured values
and status information will be collected over the specified time span and will be saved
in the meter database.
Click the "Create report" button. The Maintenance report will be generated and
displayed.
Print it and file the copy in the Manufacturer Data Report (MDR) shipped with the
meter.
The Maintenance report and the record are stored in the MEPAFLOW600 CBM
meter database and accessible via the "Meter explorer" and the "Report
Manager". Maintenance reports can be exported to an Excel file. If
Maintenance reports have been regularly collected, a Trend report can be
created (