198
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Troubleshooting
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
7.3
Generation of a Diagnosis session
If it becomes necessary to generate a Diagnosis session for remote support, follow the
procedure described below:
Start the MEPAFLOW600 CBM software and establish an online connection to the
meter (see
for all necessary preparations).
Choose "Tools / Diagnosis session" from the menu or click the "Diagnosis session" item
in the Key navigation (
)
Figure 95
"Diagnosis session" generation
Specify a file name. (The file path is set according to the program settings. If necessary,
specify a different path.)
Click the "Save" button.
MEPAFLOW600 CBM will now download the logbooks from the meter and generate a
Diagnosis session with all relevant data. The entire process usually takes about three
minutes. If the logbooks contain a lot of entries, the process may take longer.
Email the Diagnosis session file to your SICK representative for support.
"Diagnosis