Verification
Switch the meter to Operation Mode.
Verify the adjustment for a sample of preset flow rates. Deviation between reference
meter and meter under test should be within
0.1% of the reference value.
Switch the meter to Configuration Mode.
Set the pressure, temperature and meter factor back to the customized values and
write them to the meter.
Switch the flow meter to Operation Mode and switch off the power.
Remove the front cap and close the Parameter write lock (position: LOCKED) (
)
Close the front cap and switch on the power.
Documentation
Reestablish the online connection between flow meter and MEPAFLOW600 CBM.
Check in MEPAFLOW600 CBM that temperature, pressure and meter factor are at the
customized values. Compare the meter factor with the value on the main plate.
Check the "Meter status" window, ensuring that the Parameter write lock is LOCKED
and no errors or warnings are active (
)
Create and print out a parameter report. If the Calibration wizard was used, print out
the calibration report.
Sealing
Carry out the sealing (
and