Flow Calibration
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
89
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
Figure 38
"Meter calibration" wizard, step 2
Click "Add row" until you have a row for every flow test point.
Enter all flow rates Qact for the Meter Under Test and the deviation as found into the
table (
).
Select the Adjustment method you want to apply.
Click "Next".
●
The "Qact MUT" values are the flow rates of the flow calibration.
●
"Dev. as found [%]" is the difference of the test meter relative to the
reference (test stand).
●
Qact and deviation must be entered for each flow rate.
●
Only positive flows have to be entered, i.e. the reverse flow data points are
not negative flows!