5.5.2
Application data (Field setup page 2 of 8)
The pressure and temperature values which have to be entered at this page are stored as
the parameters PressureFix and TemperatureFix.
Figure 63
Field setup page 2 of 8: Application data
These values are used to correct the influence of geometric changes of the meter body due
to temperature and pressure on site (in relation to the conditions during the dry calibration
/ wet calibration). These values are also used when the FLOWSIC600 operates with the
integrated volume corrector using constant temperature and pressure correction of the
volume.
The pressure value is to be determined as described on
. The "Low flow cut
off" is usually set to 25% of Qmin.
HART
option
If the option HART
protocol was ordered, you can activate the check box
"Optional p and T reading via HARTBUS". In this case the FLOWSIC600
operates with HART® communication in Master-Mode.