5.5.7
Configuration update (Field setup page 7 of 8)
Switch the meter into the Configuration Mode.
To write the configuration and parameter settings specified on pages 2 to 6 of the
"Field setup" wizard to the meter, use the "Write to flow meter" button. The summary
field displays information about the actions just carried out (successful or unsuccessful
writing of parameters).
To reset the error volume counters and the logbooks - which is recommended after
commissioning the meter - use the "Reset at flow meter" buttons.
The time synchronization function makes it possible to write your PC time to the meter
and thus synchronize the meter with your local time settings. Be careful with this func-
tion. Read
before using it.
Switch the meter back to Operation Mode
Print a parameter report to document any changes made.
Before creating the final maintenance report according to Section
, activate the
path compensation according to
.
Figure 70
Field setup page 7 of 8: Configuration update
User Access Level: "Authorized Operator" or "Service"