Commissioning
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
145
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
5.5.9
Disconnecting from the meter and closing the session
When disconnecting from the meter, a session is stored in the MEPAFLOW600 CBM meter
database. It contains the following data:
●
a complete parameter set from the meter at disconnection
●
all parameter changes made during Field setup (entries can be viewed in the Meter
Explorer)
●
all logbook data (if downloaded)
●
the Maintenance Report created on page 8 of the Field setup
This data can be accessed later with the "Meter Explorer", even when you are not directly
connected to the meter.
To disconnect from the meter and to close the session, proceed as follows:
Go to "Connect / Disconnect" page (select "File / Connect/Disconnect" from menu).
Click "Disconnect". The "Session description" window opens.
Describe the activities carried out during the session (e.g. "Field Setup").
Click OK.