5.7
Optional advanced setup
5.7.1
Configuration and activation of User Warnings
When normal operating conditions have been reached, the User Warnings can be
configured to best suit the specific application.
To configure the User Warning limits, proceed as follows:
Use MEPAFLOW600 CBM to connect to the meter (
).
Open the "User Warnings" assistant from the main system bar by clicking the "User"
button (
Go to the "Configuration" tab (
, right side).
Use
to plan the configuration of User Warnings to
best suit the specific application.
Activate or deactivate User Warnings with the check box on the right.
Edit parameter values in the fields.
Click the "Write to meter" button.
●
The User Warnings are preconfigured when the meter is shipped from the
factory (see "Default activation state" and "Default value" in the following
tables
).
●
If there is no need to change the User Warnings, or if you are not sure about
consequences of changes, keep the values as they are or discuss with a
SICK representative.
All User Warning parameters - except for the parameter ‘Min. VOG for warnings"
- can be configured in the User Access Level "Operator" and without switching
the meter to the Configuration Mode.