O P E R A T I N G I N S T R U C T I O N S | GM32
Subject to change without notice
4
START-UP
4.14.1
OPC interface
The following data are available on the OPC interface:
Folder
Folder
Element
Data
type
Significance
Device
Status
Location
String
Installation location parameter entry. Can be set in
SOPAS ET under: Parameter - Device parameters
Failure
Bool
Device error
Maintenance
request
Maintenance request
Not Measuring
Device not in measuring operation. Set when either
maintenance, alignment mode, check cycle, zero adjust
or filter box measurement active
Check
Set when check cycle active
Uncertain
A measured value has uncertain status
Extended
A measured value has extended status
Measured
Values
Measured
Value 1
Activated
Bool
Measured value available
Name
String
Measured value identifier maximum 32 characters
Dimension
Physical unit maximum 32 characters
Value
Real
Measured value
CCycle Zero
Value
Zero point check value
CCycle Span
Value
Span check value (70%)
Failure
Bool
Measured value status error
Maintenance
request
Measured value status maintenance request
Uncertain
Measured value status uncertain
Marginal conditions for measurement (e.g. pressure,
temperature) have exceeded the allowable limit value
Extended
Measured value status extended
Marginal conditions for measurement (e.g. pressure,
temperature) are close to the allowable limit value
Measured
Value 2-16
Corresponds to Measured Value 1
Diagnosis
Lamp
Performance
Real
Lamp quality value
LED
Performance
Real
LED quality value
Start CCycle
CCycle Signal
Bool
Signal for check cycle execution
Start
Maintenance
Maintenance
Signal
Bool
Signal for maintenance mode
Disable CCycle
Disable
CCycle Signal
Bool
Signal to prevent check cycle execution
Table 8: Table of data available via the OPC interface